17
DL-00096 Rev. 7
Alarm Message
CheckVents
LowBattery
XX:Service!*
*Value:21-50
Description
Deviceisunabletomaintain
oxygenpurity.
Estimatedbatterylifeless
than17minutes.
Servicerequired.
Action
Besureairinlet/outlet
hasnotbeenblocked.If
alarm persists, contact
CustomerService.
Chargethebatterypack
by plugging in to power
supply.
Note: The message will be
automatically cleared when
plugged in to power supply.
ContactCustomer
Service.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message
CheckCannula
XX:Service*
*Value:51-70
Description
Nobreathdetectedfor5
minutes.
Servicerequired.
Action
Checkthecannulacon-
nection.Besureto
breathethroughnose.If
alarm persists, contact
CustomerService.
Note: The message will be
automatically cleared when
breathing is detected.
ContactCustomer
Service.
Medium Priority Alarms
DL-00096 rev 7.indd 17 6/1/2009 4:48:34 PM